Item # S4000, S4000 Remote Chilled Mirror Hygrometers on Michell Instruments, Inc.
Michell Instruments, Inc. - Item # S4000, S4000 Remote Chilled Mirror Hygrometers
Javascript Menu by Deluxe-Menu.com
Home
Products
By Technology
Chilled Mirror
Impedance
Calibration and Humidity Generation
Quartz Crystal Microbalance
Thermo-Paramagnetic Oxygen Sensor
Hydrocarbon Dew Point Dark Spot Principle
Capacitive Technology
Michell Zirconium Oxide Oxygen Sensor
By Industry
Aerospace
Automotive
Beverage Industry
Breathing Air
Compressed Air
Drying of Plastics
Environment
General Engineering
General Gases
Metals and Heat Treatments
Metrology
Military / Defense
Natural Gas & Petrochemicals
Pharmaceutical / Confectionery
Power Generation
Process Industries
Research & Development
Semi-Conductor / Electronics
By Category
Process Analyzers
Calibration Instruments
Portable Instruments
Dew-Point Transmitters
Chilled Mirror Instruments
Relative Humidity Products
Oxygen Analyzers
About Us
News
Exhibitions
Links
Accreditations
Support
Application Notes
Software Downloads
Support Request
Newsletter Registration
Contact Us
Search Catalog
By Keyword
By Item #
All Categories
S4000 Remote Chilled Mirror Hygrometers
|
Shipping
|
Tax
|
Terms
Shopping Cart
0 Items
Request for Quote
0 Items
All Categories
>
Chilled Mirror Hygrometers
>
S4000 Remote Chilled Mirror Hygrometers
>
Item # S4000
Download PDF
Printable Page
Email This Page
Save To Favorites
S4000 Remote Chilled Mirror Hygrometers
QUOTE
Download PDF
Email This Page
Save To Favorites
The S4000 Remote Precision Dew-point meter offers unmatched accuracy and reliability in dew-point measurements of air and gas systems and in calibration laboratories. Its powerful three stage Peltier thermoelectric heat pump gives a depression capability of more than 80ºC (176ºF) at normal laboratory temperature. The S4000 Remote can measure and control at dew points (frost points) down to -50ºC (-58ºF) from a normal room temperature, and with additional water-cooling through the integral cooling jacket can measure as low as -80ºC (-112ºF) dew point. As its name suggests, the S4000 Remote comprises a free-standing sensor and a 19" x 3 U monitor unit, connected by individual power and signal cables for the most stable and precise dew point control.
larger image
larger image
Performance
·
Electrical Input/Output
·
Output Ranges
·
Operating Conditions
·
Mechanical Specification
Performance
Measurement Technology
Chilled Mirror
Measurement Units
ºCdp, ºFdp; ºC, ºF temp; % RH, ppm
v
, ppm
w
, g/m
3
, g/kg, ppm
w
for SF
6
Measurement Range
-80 to +20ºCdp (-112 to +68ºFdp) (remote);
-80 to +85ºCdp (112 to +185ºF) (climatic sensor)
Accuracy
±0.1ºCdp (0.18ºFdp) (typical system accuracy)
Sensitivity
0.01 ºC
0.018 ºF
Response Speed
0.5ºC/sec + setting time (0.9ºF) (dew point dependent)
Repeatability
Better than 0.1ºC (0.18ºF)
Recommended Sample Flow Rate
0.1 to 0.7 L/min
0.2 to 1.5 scfh
Sensor Pressure
0.1 MPa
1 barg
14.5 psig
Electrical Input/Output
Analog Outputs
10 mV/ ºCdp, 4-20 mA
Digital Outputs
RS232: data hold, ABC status, optics alarm (logic)
Output Ranges
Display Resolution
0.01 for all units
Power Supply
90 to 265 VAC; 50-60 Hz (monitor)
Operating Conditions
Sensor Operating Temperature
-30 to +30 ºC (-22 to +86 ºF) (remote)
-30 to +90 ºC (-22 to +194 ºF) (climatic)
Mechanical Specification
Sensor Housing Type
Bench case; space frame (climatic version)
Dimensions (W X D X H)
268 x 300 x 133 mm
10.6 x 11.8 x 5.2 in
Weight
5.5 kg; 4.8 kg (climatic version)
12.1 lb; 10.6 lb (climatic version)
Maximum Cable Length
50 m
164 ft
Display
Dual 6 digit LED, with 2 digit eng unit indicators
Dual Optics Detection System
Wide band red LED with dual photo sensors
Integrated Flowmeter
0 to 1 L/min
0 to 2 scfh